AMI Designaon Capacitance
Available Part
Numbers
AMI-M12Q-1-3-B
Case
Style
A
Input/output
Style
6.3mm
Spade
Terminals
1
Current
Rat ing
3
Cx
810
L
mH
Inductance
1.036
Cy
nF pF mA
1000
Resistor
KΩ
330
Leakage
Current @
120 VAC
60Hz/250
VAC 50 Hz
0.07mA/.13mA
TIL
Inseron
Loss
001
AMI-M12Q-1-6-B A1 6 8101.83 1000 330 0.07mA/.13mA 002
AMI-M12Q-1-10-B A1 10 8101.245 1000 330 0.07mA/.13mA 003
AMI-M12Q-1-20-B B1 20 10101.76 1000 330 0.07mA/.13mA 004
Mulpurpose Medical Line Filter for
Power Line Noise Protecon
AMI-M12Q Series
Metal Enclosure Single Phase Dual Stage Filters
Electrical Schemac
AMI-M12Q
•Rated Voltage (max): 250 VAC
•Operang Frequency: 50/60 Hz
•Operang Current: 3 to 20 A
•Hi-pot Rang (one minute): Line to Ground: 2250 VDC
Line to Line: 1450 VDC
Specif icaons:
•Operang Ambient Temperature Range (at rated current Ir):
-10°C to +40°C. In an ambient temperature (Ta)
higher than +40°C the maximum operang current (Io) is calculated as follows: Io = Ir sqrt [(85-Ta)/45]