advancedenergy.com 5
COOLX1800 SERIES
ELECTRICAL SPECIFICATIONS (CONTINUED)
Leakage Currents
Parameter Conditions/Description Nom Units
AC Leakage Current Input to earth ground
Normal Condition (High Line) Mains Voltage 264 VAC/60 Hz 244 µA
Single Fault Condition (High Line) Mains Voltage 264 VAC/60 Hz 435 µA
Touch Current
Normal Condition Mains Voltage 264 VAC/60 Hz 14.2 µA
Single Fault Condition Mains Voltage 264 VAC/60 Hz 246 µA
EMC
Parameter Conditions/Description Notes
Radiated Emissions1EN 55011, EN 55022 and FCC, Class B — Compliant
Conducted Emissions1EN 55011, EN 55022 and FCC, Class B — Compliant
Power Line Harmonics EN 61000-3-2, Class A — Compliant
Voltage Flicker EN 61000-3-3 — Compliant
ESD EN 61000-4-2, level 4, 8 kV contact, 15 kV air — A
Radiated Immunity EN 61000-4-3, level 3, 10 V/m 80-2700 MHz — A
Electrical Fast Transient EN 61000-4-4, level 4, ±4 kV — A
Surge Immunity EN 61000-4-5, level 4, 2 kV DM, 4 kV CM — A
Conducted RF Immunity EN 61000-4-6, level 3, 10 Vemf 150 KHz-80 MHz — A
Power Frequency Magnetic Field EN 61000-4-8, level 4, 30 A/m — A
Voltage Dips and Interruptions EN61000-4-11 10 ms
100 ms
500 ms
A
B
B
1 Consult AE applications for system level compliance
Standards and Directives
Standard Conditions/Description
Safety Agency Approvals EN60601-1 3rd Edition, UL60601-1, CSA601, EN60950 2nd Edition, CSA C22.2 No. 60950-1
IEC/EN 60950-1, Edition 2
and all national deviations
UL 60950-1/CSA 22.2 No 60950-1, Edition 2; 5000 m (16,400 ft) altitude, 100 VAC to 240 VAC ±10%
IEC/EN 60601-1, Edition 3
and all national deviations
IEC 60601-1 (2005), EN60601-1 (2006), ANSI/AAMI ES 60601-1 (2005), CAN/CSA C22.2 No. 60601-1 (2008);
5,000 m (16,400 ft) altitude, 100 VAC to 240 VAC ±10%
IEC 62368 Edition 2 IEC 62368-1 (2014) Edition 2; 5000 m (16,400 ft) altitude, 100 VAC to 240 VAC ±10%
IEC 60601-1-2 Edition 4 IEC 60601-1-2 (2014)
Protection class Class I
ROHS EU DIRECTIVE 2015/863 RoHS compliant
REACH-171 Compliant
Conflict Materials Compliant with Conflict Free Sourcing Initiative